A simple, sensitive, and inexpensive single drop liquid‐liquid‐liquid microextraction combined with isocratic RP‐HPLC and UV detection was developed for the determination of anti‐malaria drug, chloroquine. The target compound was extracted from alkaline aqueous sample solution (adjusted to 0.5 mol/L sodium hydroxide) through a thin layer of organic solvent membrane and back‐extracted to an acidic acceptor drop (adjusted to 0.02 mol/L phosphoric acid) suspended on the tip of a 25 μL HPLC syringe in the organic layer. This syringe was also used for direct injection after extraction. The linear range was 1–200 μg/L. The LOD and LOQ were 0.3 and 1.0 μg/L, respectively. Intra‐and inter‐day precisions were less than 2.0 and 2.3%, respectively. The real samples were successfully analyzed using the proposed method. The recoveries of spiked samples were more than 94.6%.